Poly Temp Online Catalog Ultra-Low Temperature and Cryogenic Storage Freezers

IntrLogic™ Microprocessor Control

A Revco Exclusive
The IntrLogic™ microprocessor controller is the industry's most comprehensive man-agement system for temperature control, electrical power, alarm and monitoring functions. IntrLogic™ provides secure, interactive con-trol to assure the highest level of security for the stored product and refrigeration components.

Surveying ambient conditions and critical points within the refrigeration system several times each second, IntrLogic™ processes this real-time information for optimum temperature control with minimal energy consumption.

 

Setpoint Security
Key operated master switch controls main power and alarm, locks in temperature and alarm setpoints for enhanced security.

     
  A.I.M.
(Automatic Incident Monitor) Identifies any tempera-ture deviation beyond alarm setpoints, with rapid flash for a current event, slow flash if the event was self corrected. A.I.M. archives and displays at the touch of a button the most deviated temperature.
     
  Lifeguard Compressor
Protection Monitors Legaci ™ compressor performance and automatically adjusts to internal and external conditions, extending component life.
     
  Extreme Ambient Alert
Warns if room temperature is too warm due to seasonal air conditioning or other factors which may affect freezer performance.
     
  Automatic Voltage Boost
Compensates for low voltage and brown-out conditions; IntrLogic™ adds an on-board power monitoring function with digital read-out of incoming line voltage at the push of a button.
     
  Data Entry
Touchpad data entry and function selectors feature a simple input protocol designed to avoid human error.

 

     

     
    IntrLogic™ features a standard RS-232 data port with system software for quick PC installation. Remote data logging functions include set-point, high and low limit and power supply monitoring to advise of temperature deviation, voltage boost, surge intercept, low battery and prompt to clean filter. An automatic status report is adjustable in intervals of 5 seconds to 2 hours. Click Here for a PDF version of the schematic above (813k).
Back to Main ULT and Cryogenic Freezer Page
Back to top.
Poly Temp Main Navigation