Ultra-Low
Temperature and Cryogenic Storage Freezers |
 |
Ultima
II Series
Cryogenic
Storage Freezers


Model
ULT-10140-9, 10.3 cu.ft. capacity (292 liter), -140°C, shown with
optional temperature recorder. |
|
| Refrigeration |
Control |
     |
     |
| Orbital |
IntrLogic |
Choose
from two chest models, two cryogenic operating temperatures. Model
ULT-7150-9, -150°C, shown with optional temperature recorder.
|
| |
|
|
Ultima
II Series Cryogenic Operating Temperature
-150°C,
Model ULT-7150-9
-140°C,
Model ULT-10140-9
Air-cooled
refrigeration systems are engineered to maintain design temperatures
in warm ambient conditions:
Model ULT-7150-9, +30°C (+86°F);
Model ULT-10140-9, +32°C (+90°F)
Cryogenic
Temperatures Without Liquid Nitrogen
Revco
Ultima II cryogenic freezers are designed for long-term storage of biologicals
at uniform temperatures of -140°C and -150°C, safely below the -130°C
glass transition temperature of water
Cell
viability is improved because biologically relevant, thermally driven
reactions are not known to occur
Safer,
More Convenient Than LN2
Revco
cryogenic freezers include all features, control and performance attributes
of the Ultima II Series freezers, plus many important advantages not
available when using LN2
Stable
cryogenic temperature of -140°C and -150°C without LN2 as a cooling
medium
Patented,
single compressor, orbital refrigeration system
Top-to-bottom
temperature uniformity dramatically improved when compared to LN2 vapor
Lower
operating costs than LN2 freezers of similar capacity
100%
of interior space is usable, with stored samples maintained below critical
temperature
Ample
interior volume with capacity of up to 21,600 samples in Model ULT-10140
and up to 14,400 samples in Model ULT-7150 (2 ml vials in boxes)
Uses
standard Revco inventory racks and fiberboard boxes; see Accessories
Refrigeration
System

Exclusive mixed refrigerant autocascade with a single, orbital refrigeration
compressor, air-cooled
CFC-free
refrigerants
Downfeed
evaporator for most efficient refrigerant flow
High
capacity air-cooled condenser with dual condenser fans, aerodynamically
shaped fan blades
Washable
condenser filter keeps fins free of dust to maintain peak cooling efficiency;
no tools required for removal

Comparison of temperature distribution in a liquid nitrogen freezer
(vapor phase) and mechanically refrigerated cryogenic freezer (i.e.
Revco Ultima II Model ULT 10140-9). Graph represents temperatures (°C)
at relative locations within the chamber. Data suggests that 100% of
the Ultima II storage space maintains uniform storage temperatures safely
below -130°C, while temperature in LN2 vapor system is dependent on
storage location.
Control
System
Exclusive
IntrLogic microprocessor control system with touchpad data entry
and digital display of all functions
Setpoint
security system with key-operated switch for main power and alarm locks
in temperature and alarm setpoints
Independent
operating temperature and high/low limit alarm functions with factory-set
defaults for quick start-up, all setpoints adjustable in 1°C increments

Ultima II
IntrLogic™ microprocessor controls monitor all functions for safe,
dependable and economical operation; shown with optional LN2 back-up
system. |
| |
Electrical
System

Automatic voltage boost compensates for low voltage and brown-out conditions
On-board
power monitoring with digital read-out of incoming line voltage
Built-in
surge suppressor minimizes effect of voltage spikes from incoming electrical
service
Alarm
System
A.I.M. (Automatic Incident Monitor) identifies any temperature deviation
beyond alarm setpoints, with rapid flash for current event, slow flash
if event was self corrected; A.I.M. archives most deviated temperature
Adjustable
safety alarm with automatic, continuous charge battery back-up provides
full alarm function in case of power failure
Power
failure or temperature deviation triggers audible and visual warning
Digital
battery indicator displays battery capacity; flashing low battery indicator
advises when to replace battery
Touchpad
alarm test function physically warms internal probe to simulate alarm
condition
RS-232
data port with system software package
Exterior
alarm contacts for connection to remote monitoring system
Monitoring
System
Lifeguard
compressor protection monitors performance and automatically adjusts
to internal and external conditions, extending compressor life
Extreme
ambient alert warns if room temperature may affect performance
Voltage
boost indicator advises of low source voltage below safe parameters
and indicates voltage boost system is activated
Surge
suppressor indicator advises to check intercept fuse
Clean
filter indicator light advises when to remove and wash condenser filter
Chest
Cabinet Construction
Low
profile design for easier access to inventory
Heavy-duty
counter-balanced lid with integrated keyed lid lock for security; 3.1
cu.ft. (88 liter) model includes locking provision only
Double
seal lid gaskets to minimize frost build-up
Independent
sub-lids provide additional protection
Minimum
5" (12.7 cm) foamed-in-place urethane insulation, closed cell,
CFC-free, to protect product against high ambient conditions
Heavy-gauge
steel cabinet with high-impact powder paint finish
Rounded
interior corners for easy cleaning
Access
port, 1" (2.54 cm) diameter, with cap
Recessed
heavy-duty swivel casters
Inventory
System and Accessories
Optional
inventory components including racks, boxes and dividers may be ordered
separately or as full-load inventory systems
See
Accessories section for information on optional temperature recorders,
free-standing LN2 back-up systems, inventory components, validation
testing, calibration certificates and associated products
|